13-3. Connecting the Supply Hoses
q
Connect the supply hoses to the 25 liter replenisher tank.
Yellow
• • • • • • • • • • • •DEV Supply Hose.
Red • • • • • • • • • • • • • •FIX Supply Hose.
w
Use SNP-12-HSO to clamp the hoses to the tank.
NOTE :
Do not extend any of the hoses more than 5 meters in length. If these 12 mm diameter hoses
are extended longer than 5 meters, they tend to bend and collapse.
13-4. Connecting the Drainage Hoses
q
Connect the drainage hoses to the 25 liter replenisher tank.
Yellow
• • • • • • • • • • • •DEV Drainage Hose
Red • • • • • • • • • • • • • •FIX Drainage Hose
Transparent(Thick) • • • •Water Drainage Hose.
Transparent(Thin) • • • • •Overflow Drainage Hose
w String the chemical and water drainage hoses so that no liquid will accumulate in the hoses
between the main unit and the drainage tank or the water drain.
13-5. Power Source and Wiring
q
Changing the terminal Board to Meet Voltage Requirements
Follow these procedure in changing the connection to terminal board
TB1 located on the top of the electrical components unit.
UL Version
1
2
£ £ £ £ £ £ £ £ £ £ £ £ £ £
£ £ £ £ £ £ £ £ £ £ £ £ £ £
CE Version
1
2
£ £ £ £ £ £ £ £ £ £ £ £ £ £
£ £ £ £ £ £ £ £ £ £ £ £ £ £
INSTALLATION INSTRUCTIONS AND PRECAUTIONS
3
4
5
6
7
8
9
10 11 12 13 14
g g
Black Wire
w q
q For AC 120V outlets, connect terminals #5 and #8.
w For AC 115V outlets, connect terminals #5 and #7.
3
4
5
6
7
8
9
10 11 12 13 14
g g
g
Black Wire
w q
e
q For AC 240V outlets, connect terminals #5 and #9.
w For AC 230V outlets, connect terminals #5 and #8.
e For AC 220V outlets, connect terminals #5 and #7.