Further Information On Device Configuration In Program Mode; Standard Setup - KROHNE OPTIWAVE 3500 C Handbook

80 ghz radar (fmcw) level transmitter for liquids with hygienic requirements
Table of Contents

Advertisement

OPERATION
6

6.4 Further information on device configuration in Program mode

6.4.1 Standard setup

Use this procedure (menu item A4.1 Standard Setup) to change the length unit, tank type, tank
height, current output variable, 0% range, 100% range, current output range and error function.
Values and parameters that can be changed are shown between the « ... » marks in the
illustrations that follow. Push the keypad buttons in the correct sequence:
CAUTION!
Make sure that you do this procedure before you use the device. The settings in this procedure
have an effect on the performance of the device.
INFORMATION!
Current output and LCD display settings
When you do the Standard setup procedure, the measurement value on the first measurement
page of the LCD display (in Normal mode) automatically uses settings with the same parameters
and values as the current output settings (measurement variable, 0% range and 100% range).
The first measurement page shows "one value and bar" by default. If you must show one or more
measurement values that are different to the transmitted output signal, change the settings in
the C6.4 1st Meas. Page and C6.5 2nd Meas. Page menus. For more data, refer to Function
description on page 70 – table C. Full Setup menu (C6 Display).
Procedure
Screen
90
Steps
2 × [>], 2 × [ ▼ ] and [>].
If it is the default password: 2 × [>],
5 × [ ▲ ], [>] and 8 × [ ▲ ]
[ ↲ ] to confirm.
[ ▼ ] and 2 × [>]
[ ▲ ] or [ ▼ ] for the selection of the
length unit (m, Cst., in, ft, mm or
cm).
[ ↲ ] to confirm.
[ ▲ ] or [ ▼ ] for the selection of the
tank type (Storage, Stilling Well,
Agitator or Process).
[ ↲ ] to confirm.
www.krohne.com
OPTIWAVE 3500 C
Description
Default screen.
Enter Program mode and go to menu item
A3 Login.
Enter the password used at this time for
the "Expert" access level. If it is the default
password, enter "0058".
Push these buttons to start the Standard
Setup procedure.
Length unit. Make a selection from the list
of parameters.
Tank type. Make a selection from the list of
parameters. If the surface of the product is
flat, select "Storage". If the surface of the
product is disturbed, select "Process". If
the surface of the product is agitated with
vortexes and foam, select "Agitator".
02/2023 - 4005813604 - MA OPTIWAVE 3500 R04 en