Configuration
January 2023
15. Click the Service Tools → Routine Maintenance tab. Click Analog Output 1
16. Click the Service Tools → Zero Calibration tab. See AMS Device Manager -
17. After you have changed and written the configuration changes to the meter
88
Figure 4-19: AMS Device Manager - Service Tools Trends
Primary and Secondary variables display real-time uncorrected
volume flow rate trends. The third and fourth variables charts display
trends for temperature and pressure.
Trim to perform a digital to analog trim adjustment of the first milliampere
output. The 4 mA and 20 mA output current values should equal the plant's
standard values.
a) Click Yes to confirm the configuration changes. Repeat this step to
trim Analog Output 2 current.
b) Click Apply to write the output trim values to the meter.
c) Click OK to navigate back to the Service Tools page.
Guided Setup
Step 7
to configure the zero flow parameters.
do the following:
a) Enable the Write Protect switch on the CPU Module to protect the
meter's configuration.
b) Replace the end cap and if required, apply security seals through the
endcap holes and through the hex head bolts that secure the Bracket/
Cover to the Base enclosure.
Note
The next time you connect to the device using MeterLink, the Monitor
page displays a Meter status alarm that the configuration has
Installation manual
00825-0100-3812
Rosemount 3812 Liquid Ultrasonic Flow Meter