1. Before the pipette controller is activated for the first time, the batteries should be
charged.
2. FASTPETTE PRO should not be left uncharged for a longer period of time.
3. In the case of a planned longer break in the work of the pipette controller, it is recommended
to take the batteries out of the pipette controller.
9 - MAINTENANCE
Cleaning
The FASTPETTE PRO does not require any maintenance. The external parts of the pipette
controller may be cleaned with a swab moistened with isopropyl alcohol.
The nosepiece and the pipette holder may be autoclaved at 121 °C for 20 mi nutes. After auto-
claving, dry the pipette holder. The filter included in the set may be sterilised by autoclaving
at 121°C for not more than 15 minutes.
Storage
The FASTPETTE PRO shall be stored in a dry place. The allowable storage temperature: -20 °C
to +50 °C.
During breaks in the work the pipette controller shall be placed on a charging stand in hori-
zontal position (fig. 7A) or in vertical position (fig. 7B). The charging stand may be fitted on
a wall (fig. 7B) with a fixing set included with the pipette controller.
CAUTION:
Do not store the pipette controller with a filled pipette.
10 - COMPONENTS
The FASTPETTE PRO set is supplied with the following components:
• Pipette controller
• Charger
• Charging stand
• Membrane filter 0.2 μm
• Battery – 3 pcs.
• Instruction manual
7
11 - ORDERING INFORMATION
The FASTPETTE PRO comes with a universal charger and a set of adapters in different versions:
EU, US, UK and Australia. Choose your country used adapter and connect to the housing.
To mount the adapter, it should be inserted into the slots of the housing in the direction of the
arrow (1), until you hear a click engagement.
To remove or change the adapter, simply press the "PUSH" button in the direction of the
arrow (2), then holding the button down, remove the adapter in the direction of the arrow (3).
Catalogue no. P2002
12 - SPARE PARTS
Item
Part name
in fig. 1
F
Nosepiece
G
Pipette holder
H
Membrane filter 0.2 μm
Membrane filter 0.45 μm
J
Connector gasket
L
Battery - NiMH
M
Charging stand
N
Charger 9V: EU, US, UK, AU
13 - PRODUCT DISPOSAL
According to Directive 2012/19/UE of the European Parliament and of the Council of
4 July 2012 on waste electrical and electronic equipment as amended,
pipette controller FASTPETTE PRO is marked with the crossed-out
wheeled bin and must not be disposed of with domestic waste.
In accordance with the requirements of Directive 2006/66/EC of
6 September 2006 on batteries and accumulators and waste batteries
and accumulators, batteries must be disposed of in accordance with
national regulations.
The crossed out wheeled bin symbol is printed in the product instruction manual and on the
packaging. In order to recycle the used product, send it to the nearest service centre or to
the manufacturer. For contact details please go to www.labnetinternational.com
All rights reserved. Product described in this manual is subject to availability and technical modification. Errors excepted.
LABNET reserves the right to improve, enhance or otherwise modify its products without prior notification.
© 2014 LABNET
FPPR/11/2014/I
Number of
Catalogue no.
pie ces in pack
P29052
1
P29054
1
P9023
1
P9024
1
P9025
1
P29066
3
P29055
1
P9010
1
8
Need help?
Do you have a question about the FASTPETTE PRO and is the answer not in the manual?
Questions and answers