Additional Notes

The analyzer shared data directory is in the LGR workgroup. If it is not visible, browse for it in
the Windows Network Neighborhood. The computer name will be XXXXXXXXXXXX, where
XXXXXXXXXXXX is the analyzer serial number.
The current data file of the analyzer can be open while measurement is in progress without
interrupting the analyzer operation. The current data file is updated after every fourth KB, so a
new data file appears empty until enough data is collected to be written to the disk.
If a LAN is not available, plug the analyzer into a standalone broadband router (example:
Netgear Model RP614) to enable the analyzer to obtain a DHCP address from the router when
the analyzer is started. Then, plug any Windows computer into the same broadband router to
access the data directory.
A crossover Ethernet cable does NOT allow an external computer to access the shared data
directory, because the analyzer will not obtain a DHCP address on initialization and will shut
down its Ethernet interface.
It is possible to access the shared analyzer data directory from operating systems other than
Windows. The analyzer uses a Samba server to share the data directory, which could be
accessed by any appropriate Samba client application.