ABB LGR-ICOS™ GLA232 Series Methane/Ethane Analyzer User Manual
About
The About section displays analyzer specific information, such as the:
Current software build date
Code version
IP address
Analyzer serial number
Laser Adjust Tab
Use the Laser Adjust tab to manually adjust the laser's wavelength to compensate for any
cumulative drift. (Figure 22)
Laser adjustment may be required for the following reasons:
The laser's wavelength has drifted beyond the target range of the analyzer.
The analyzer is operated outside the recommended temperature range.
Figure 22 shows the offset between absorption peaks and target lines for laser A (top plot) and
laser B (bottom plot). Both lasers are reasonably well adjusted.
ABB PROPRIETARY INFORMATION
Figure 22: Example of Laser Adjust Tab
36