Option ST1 and ST2 – IQ Streaming
A22 Ultra-Wideband Analog IF ODI Cable Disconnect – Option R40
1. Referring to
Figure
assembly (1), carefully disconnect the Optical Data Interface (ODI) cable coming from the A22
assembly (2) from the center connector block (4) by sliding the blue locking tab away from
the connection.
Figure 6
A22 Ultra-Wideband Analog IF ODI Cable Disconnect – Option R40
2. Carefully disconnect the Optical Data Interface (ODI) cable coming from the A21 assembly (3)
from the center connector block (4) by sliding the blue locking tab away from the connection.
3. Remove the two panhead screws (5) along with the two cable clamps.
14
6, if the instrument has the option R40 A22 Ultra-Wideband Analog IF
Great care needs to be taken when installing, removing, and handling any of the Optical Data
Interface cables as they can be easily damaged if they are bent too tight or pinched.
Installation Note N9042-90011