Footprint/ sample installation/ weights
DuraMax without base
60
670
Front and side view/ M 1:30/ dimensions in millimeters
For the installation of DuraMax, you can also use a sturdy table as a base. It should be at least 700 mm high and
must be designed to bear a load of ≥450 kg. (Provided by customer).
Other dimensions (e.g. distance to the wall) and required connections can be seen in the drawing on the next page.
Weight
CMM
Workpiece, max.
Floor load
Mean pressure on the support surfaces.
Please check the corresponding support surfaces.
Warning for a CMM on a table
If the CMM is on a table, the CMM can slip and fall off.
WARNING
There is a risk of injury if the CMM falls off due to vibrations.
Crushing of body parts
Irreparable damage to the CMM.
Provide suitable anti-slip protection (e.g. a collar mounted around the storage surface or depres-
sions in which the CMM feet can be placed).
14
1080
DuraMax Installation site requirements
835
1170
1360
Value
[kg]
350
[kg]
100