Hp 8341A/B; Introduction; Selecting The Type Of Testing - HP 8340B Manual

Synthesized sweepers
Hide thumbs Also See for 8340B:
Table of Contents

Advertisement

Chapter
7.
Testing an
HP
8340A/B
or
HP 8341A/B
7.0
Introduction
Whether
you are testing an
HP
8340A/B
or
HP
8341A/B., the tests which may be
run
and the
flow of the test
software are
the
same-
Since this
is
the case,
this
chapter
will
discuss
the flow
of the test software from the
8340 FAMILY
TEST SYSTEM
menu
to
the test
menu
(either the
Operation Verification menu
or the Attenuator Calibration menu) This
chapter
will
not differ¬
entiate
between
testing an HP
8340A/B
or
HP 8341A/B Once
you are ready to run
a
specific
test, you should
proceed
to the chapter which discusses the test
you
wish
to
run
7.1
Selecting the Type of Testing
After
you
have selected the appropriate softkey, Test8340A, Test8341A, TestH02 (HP
8340A
Option H02),
Test8340B, orTest8341B,
you
will be given
a
choice of performing an
Operation
Verification (OpVer) or Attenuator Calibration (Cal) For either type of testing you select, the
program
will
proceed as
follows.
1.
The computer
will
tell you the
model number and serial number
it
read from
the Instru¬
ment's Calibration Constants and
ask you to
verify that they are correct-
If
the
number
obtained
is
incorrect you
will
need
to
verify the instrument's calibration data
using the
hard copy of the calibration
data
located underneath the
DUT's top cover. Otherwise,
select YES
2.
The computer
will
now read the
rest of the
Instrument's Calibration Constants
If
the
computer now
offers the
Menu
softkey, go directly to
step
3
CALIBRATION CONSTANT DISCREPANCIES
If
there
is
a
Cal Constants data
file for
this instrument on the
data
disc, and the
file
does
not match the Cal
Constants
in
the instrument,
the computer
will
display the following:
OUT CAL #Ca/
Constant Name
=
### BUT FILE
=
###
[OUT OK]
[NEITHER]
[FILE OK]
The most common reason for
this
screen to be displayed
is
that
someone has pressed
[SHIFT] [PEAK]
on the synthesizer- This
function causes the synthesizer
to automatically
optimize
its
RF output power and to update the
YTM
GAIN,
YTM OFFSET,
and the YTM BX
GAIN Cal Constants.
If
the Cal Constant
being displayed
Is
one
of
those
listed
above, and
the
synthesizer does not have
a
front panel failure annunciator on,
it is
likely that the
Cal Constant
In
the
DUT
is
OK.
In
any case,
it
Is
up to
the operator to determine which of
the Cal
Constants
is
valid
(one method
is
by
examining the printed copy of
the Cal
Constants located underneath
the synthesizer's top cover) or to assume that the data disk (file)
Is
OK.
Press the appropriate softkey
in
response
to the
prompt.
If
NEITHER
is
selected, you
will
need
to
enter the correct value for the Cal Constant After
all
Cal Constant discrepancies are
taken care
of,
the program
will
return
to
normal operation
HP 8340A/B and HP 8341A/B
Testing an
HP 8340A/B
or
HP 8341A/B
21