The softkeys provided for
this
test are as follows:
ADJUST
MEASURE
SPEC
Viewdat
Hardcpy
JANITOR
-
Does everything-
It
adjusts the Calibration Constants; verifies
power
accuracy at every attenuator step;
it
also sums the attenuator perfor¬
mance data
with the output
power accuracy data
and
compares
the
results to instrument specifications After
this
key
is
selected, press
the ALL softkey to adjust and measure
all
of the attenuator steps.
If
only
one step
is
selected, no data
will
be stored
-
This
is
a
subset of
the
ADJUST
test.
MEASURE
verifies the low
output
power accuracy
at
every attenuator
step;
it
also sums
the
attenuator performance data
with the output
power accuracy data
and compares the
results
to
instrument specifications. After
this key
is
selected, press the
ALL softkey
to
measure
all
the
attenuator
steps and store the
data.
If
only one step
is
selected, no data
will be
stored
-
This
is
a
subset of
the
above
tests.
SPEC
simply
takes
existing
attenuator performance
data
and sums
it
with
output power
accuracy data The
result
is
compared
to
instrument specifications
-
Allows you to view the test result data on the
computer's CRT,
After
this
key
is
pressed,
you
are
given
a
choice of viewing the Individual
attenuator step data or viewing
all
the
data.
-
Same
as the
Viewdata
softkey except
the
selected data
is
output
to
the printer
-
Returns you
to the test
menu-
Select
the
desired softkey and respond
to the
prompts
provided by the program-
NOTE
The test setup
for the Adjust and Measure portions of the RF
Attenuator
Calibration
is
extremely
critical. The tests and adjustments are
per¬
formed at
low RF output
power levels (down
to
—90 dBm). Use of
a
defective cable
or
adapter,
or having
a
loose
cable
connection
in
the test
setup may cause
a
misadjustment of the Instrument
or failure
in
the low
output power accuracy test.
If
the
test results indicate that the
instru¬
ment failed
low output power accuracy,
check the
test setup and re-run
the
selected test
(either Adjust
or
Measure).
NOTE
The program
will
repeat the "Set
IF
attenuator" statement for each
mea¬
surement, even though
the proper
attenuation has already been set.
HP 8340A/B and HP 8341A/B
RF Attenuator Calibration