4-4.
SWEPT FREQUENCY ACCURACY
Specification
Table 4-7.
Swept Frequency Accuracy Specifications
Center Frequency/Sweep Width Mode (CF/AF)
Readout
Accuracy with respect to
sweep
out voltage
(sweep
time
>
100
milliseconds):
AF
^
n
x
5
MHz:
±
1% of indicated sweep width (AF)
±
time base accuracy*
AF
>
n
x
5
MHz
to
<
300 MHz:
±
2%
of indicated
sweep
width (AF)
AF
s
300 MHz:
±
1%
of indicated
sweep
width (AF), or
±
50 MHz, whichever
is
less
Where
n
=
harmonic multiplication number.
Refer
to Frequency Ranges and Bandswitch Points
description
in
the specifications table
in
Section
I.
*
Time Base
affects
Center Frequency accuracy
only, not
Sweep
Width Accuracy.
Start/Stop Mode
Readout Accuracy
with respect to
sweep out
voltage (sweep time
>
100
milliseconds):
Same as Center Frequency/Sweep
Width Mode.
Description
The HP 8340B/41B (DUT) RF
output
is
connected to the HP 8566B Spectrum Analyzer input. The
spectrum analyzer
is
set
for
zero Hz span
at
a
CW frequency
within the
DUT swept
frequency range.
The spectrum analyzer VIDEO
OUT
(applied to the oscilloscope vertical input) will have
a
response
as
the
DUT
output passes through the
frequency that
the spectrum analyzer
is
tuned
to.
The selected
spectrum analyzer bandwidth
filter will
determine the shape of the response.
The HP 8340B/41B SWEEP OUT,
in
series
with
a
power
supply, is
applied to an oscilloscope's
horizontal input. The input to
the
oscilloscope
is
clamped
at
±
0.7V by
two diodes
to
prevent overdriv¬
ing the
oscilloscope's
input.
The oscilloscope
is
calibrated by setting the
DUT
and the spectrum
analyzer
to
the same CW frequency (e.g., 20% of
Band).
To
set the
DUT
to have
a
SWEEP OUT
voltage and at the same time
a
phase locked CW frequency that
is
proportional
to
a
percentage
of
a
swept frequency
band, the
START
and
STOP
frequencies are selected,
MANUAL SWEEP
is
selected,
and
a
frequency equivalent
to the desired
percentage of band
is
selected (e.g.,
START
3
GHz,
STOP
5
GHz, MANUAL 3.4 GHz, for 20%). The power supply
is
adjusted for
a
DVM
indication of
zero
volts and
the
oscilloscope horizontal
position control
is
used to position the dot to the center graticule
line.
With
the
oscilloscope horizontal
sensitivity set
to 0.05 V/Division, the
CRT
horizontal axis
is
now calibrated
to
approximately 0.5% of the swept frequency range per
division.
The spectrum analyzer VIDEO
OUT
voltage to the
oscilloscope
is
then maximum
and
the scope
vertical position
is
adjusted
to
place
the
dot
near
the top of the
CRT.
The HP 8340B/41B
can
now be set to sweep
any
frequency range and the oscilloscope center
grati¬
cule line
will
represent the
calibrated percentage (e.g.,
20%)
of the sweep range.
The
spectrum
analyzer
center frequency
is
set
to
a
frequency that
is
the desired percentage (e.g., 20%) of the DUT
swept frequency range.
If
the
DUT swept frequency accuracy
is
perfect, the oscilloscope trace
will be
the
response of
the
spectrum analyzer's bandwidth
filter,
centered
on
the
CRT.
If
the trace
is
not
centered, the spectrum analyzer center frequency
is
adjusted
to position the
response to the center of
the oscilloscope
CRT. The amount
of
spectrum analyzer frequency change
is
the
DUT Swept
Fre¬
quency Accuracy
error.
HP 8340B/41B
Performance Tests
4-21