Spurious Signals Test - HP 8340B Manual

Synthesized sweepers
Hide thumbs Also See for 8340B:
Table of Contents

Advertisement

4-7.
SPURIOUS SIGNALS TEST (10 MHz to 22 GHz
-
HP
8340B
only)
(10 MHz
to 20 GHz
-
HP 8341
B)
Specification
Table 4-16.
Spurious Signals Test Specifications
SPURIOUS SIGNALS
(Expressed
in
dB relative to the
carrier level (dBc) at ALC
level
of
OdBm)
Harmonics
(up to
max. output freq.)
Subharmonics and multiples
thereof
(up to max. output freq.)
Non-harmonically related
spurious
(CW
and
Manual
Sweep mode
only)
Power
line
related and spurious due to fan rotation within
5
Hz
below
line
frequency, and multiples
thereof
(CW mode
only,
all
power
levels)
STANDARD
INSTRUMENT
Offset
<300
Hz
from carrier
Offset 300 Hz to
1
kHz from
carrier
Offset
>
1
kHz
from carrier
Bands and Approximate Frequency Ranges
(GHz)
(See Frequency Ranges
and Bandswitch Points for complete description)
Band
0
0.01 to
<2.3
<-35
-
<-50
<-50
<-60
<-65
Band
1
2.3 to
<7.0
<-35
-
<-70
<-50
<-60
<-65
Band
2
T.Oto
<13.5
<-35
<-25
<-64
<-44
<-54
<-59
Band
3
13.5to <20.0
<-35
<-25
<-60
<-40
<-50
<-55
Band 4'
20.0 to 26.5
<-35
<-20
<-58
<-38
<-48
<-53
1
Band
4
applies
to HP
8340B
only
Description
The HP 8340B/41B RF output signal
is
displayed on
a
spectrum analyzer to
verify
that
the
harmonic and non-
harmonic spurious
signals are at or below the specified level.
Equipment
Spectrum Analyzer
...................
Adapters
TYPE
N
(m) to APC 3.5
(f)
............
(2 required
if
DUT
is
HP
8341
B)
(1 required
if
DUT
is
HP 8340B)
APC 3.5 (f)
to
APC 3.5 (f)
............
(1 required
if
DUT
is
HP 8340B)
Cable (Semi-Rigid
SMA
male
to
SMA
male)
.......
HP 8566B
HP P/N 1250-1744
...
HP P/N
5061-5311
HP
P/N 08340-20124
HP 8340B/41B
Performance
Tests

Advertisement

Table of Contents
loading

This manual is also suitable for:

8341b

Table of Contents