5-15.
ALC
ADJUSTMENTS (Cont'd)
32. Adjust
the
power
supply for
+0.3
Vdc. The power meter
indication should
be between 2.18 to
2.38 dB.
If
the indication
is
not
in
this
range, adjust A26R91 (AM GAIN, Figure 5-51) for
a
power
meter indication within the given range.
A26
a
DD
iimiminmmimir
Figure 5-51.
A26
Linear Modulator Component Location
Diagram
33. Adjust the
power supply
for —0.3 Vdc. The
power meter
indication should be
between
—2.91 and
-3.29
dB.
If
the indication
is
not
in
this
range, adjust A26R91 (AM GAIN) for
a
power meter
indication within the
given range.
34.
Repeat
steps 32 and 33
until
the power
level
requirements are met for each step.
NOTE
If
adjustments of the calibration constants were made
in
this
procedure,
the data thus stored
in
the "working memory area"
should
be copied to
the "protected memory area"
by
pressing the following key sequence:
[SHIFT]
[MHz] [1] [4] [Hz] [SHIFT] [kHz] [5] [3] [4] [9] [Hz].
Wait
for
"CALIBRATION RESTORED"
to be displayed
in
the ENTRY DISPLAY and
then press [INSTR PRESET].
HP 8340B/41B