4-5.
MAXIMUM
LEVELED OUTPUT POWER AND POWER ACCURACY
TEST (Cont'd)
HP
8340B/41B
UNDER TEST
OSCILLOSCOPE
Figure 4-5.
Maximum
Leveled
Output
Power and Power Accuracy Test Setup
Procedure
MAXIMUM LEVELED POWER
1.
Connect the equipment as shown
in
Figure
4-5. Connect the
10
dB pad and crystal detector to the
DUT
RF output and connect the detector output
to
the oscilloscope channel
A
input. Allow the
equipment to warm up
for
one hour.
2.
Press
[INSTR PRESET].
Press
[PEAK]
on. Press
[STOP FREQUENCY]
and enter
[2]
[.] [3]
[GHz].
Adjust the oscilloscope to view the swept output
of the crystal
detector.
In
Band
0
(10
MHz
to 2.3 GHz), the crystal detector may pass
a
portion of the
RF
signal. This
RF
feedthrough may
be visible on
the low frequency
portion of
the oscilloscope trace.
Also, as
the power
in
Band
0 is
increased
toward
maximum leveled, the level of the harmonics
may increase and cause
a
perturbation
in
the swept response. This
will
cause the maximum
power
indication on the
oscilloscope to increase without
causing the UN
LEVELED
indicator to
light.
Therefore, flatness
is
only specified at
^
+10
dBm
in
Band 0.
3.
Press the [POWER LEVEL]
key and, using
the
front
panel rotary knob, increase the power
level
until
the
UNLEVELED
Indicator
comes
on. Slowly
decrease the power level
until
the
UNLEVELED
indicator Just goes off.
Note the ENTRY DISPLAY power
indication.
Performance Tests