HP 8340B Manual page 25

Synthesized sweepers
Hide thumbs Also See for 8340B:
Table of Contents

Advertisement

4-3.
SWEEP TIME ACCURACY (Cont'd)
3.
Set
up the 5316A as follows:
TI
A—B
.......................
GATE
TIME
...................
LEVEL/SENSE (CHANNEL
A
and
B)
TRIGGER (CHANNEL
A
and B)
...
AC/DC (CHANNEL
A
and
B)
......
ATTN (CHANNEL
A
and
B)
.......
FILTER
......................
SEP/COM
A
...................
Channel
A
Triggering
...........
Channel
B
Triggering
............
..........
IN
........
MIN
......
+MAX
......
LEVEL
.........
DC
..........
X1
......
NORM
.....
COM
A
Leading Edge
.
Trailing
Edge
NOTE
To ensure that an accurate sweep time indication
is
obtained from
the
HP
5316A
in
step
4,
allow the DUT to complete at least three
sweeps.
4.
On the HP
8340B/41B, press [SWEEP
TIME] [1] [0] [msec].
Verify that the
sweep time
displayed by the HP 5316A
is
within the
limits
shown
in
Table 4-6.
5.
Repeat step
4
for values shown
in
Table 4-6.
Table 4-6.
Sweep
Time Accuracy
Limits
Selected
10ms
100 ms
1
s
10s
50s
Lower
Limit
9.5
ms
95 ms
.95s
9.5s
47.5s
Upper Limit
10.5 ms
105 ms
1.05s
10.5s
52.5s
Performance Tests
HP 8340B/41B

Advertisement

Table of Contents
loading

This manual is also suitable for:

8341b

Table of Contents