11.3 BAKE TEMPERATURE SET POINT IS NOT ACHIEVED:

Bake Temperature set point is not set to the value you want:
Confirm Bake Temperature set point value.
o
Bake Time has completed or the Bake time was not set to the expected value:
Confirm Bake time value.
o
The product in the cabinet is "High-Mass" and may take longer to process.
o
You may not have purchased the SmartDRY™ SD-10SB cabinet with SmartBAKE™. Normally the Con-
troller will not allow you to program the SmartBAKE™ cycle unless you purchased this cabinet. How-
ever, if SmartBAKE™ is in your display menu and you did not purchase it, contact ECD and we can cor-
rect that error in configuration.
The Bake heater may be damaged.
The RH/Temperature Probe is damaged:
Replace the probe following the procedure outlined in the Maintenance section: RH/Temp Probe.
o
SD Series Dry Cabinets Setup and Operation Manual | A55-7379-00 Rev: 2.0.0
Page 74