11.0 SERVICE AND TROUBLESHOOTING

11.1 MAIN DISPLAY IS BLACK:

The main display should power on within 30 seconds of apply power. If not, try the following:
There is no power at the power plug:
Confirm the power plug is energized.
o
The controller did not correctly reset:
Unplug the power, wait 10 seconds or so and reconnect to the cord to the power plug.
o
The circuit protection fuse has "blown" and need to be replaced:
Always disconnect the cabinet from the power cord from the power supply before replacing the
o
protection fuse.
Remove the control chassis top cover screws located on the cover's back
o
Slide the cover back about 20mm (1") and lift off the top. Be careful to prevent the cover from
o
falling inside the control chassis.
Find the fuse on the Power Board and remove insulating cover and the old fuse. Replace the fuse
o
with a 3AG Type Slo-Blo 10Amp for 120VAC supply voltage, 5Amp for 240VAC supply voltage as
noted on the Power board. Refer to Section
Replace the insulating cover over the fuse
o
Replace the Top cover and the fastening hardware BEFORE connecting the power cord to the
o
power supply.
SD Series Dry Cabinets Setup and Operation Manual | A55-7379-00 Rev: 2.0.0
Appendix A: Parts List
for reordering information.
Page 72