Eighteeth UltraMint Pro User Manual page 24

Ultrasonic scaler
Table of Contents

Advertisement

Functional
Testing,
Maintenance:
Packaging:
Sterilization:
Storage:
6
Cleaning, Disinfection and Sterilization
Visual inspection for cleanliness of the instruments and
reassembling.
Functional
instructions of use. If necessary, perform reprocessing
process again until the instrument is visibly clean.
Before packaging and autoclaving, make sure that the
components have been maintained according to the
manufacturer's instruction.
Pack the instruments in an appropriate packaging
material for sterilization.
 Check the validity period of pouch given by the
manufacturer to determine the shelf life.
 Use pouches which resist to a temperature up to
141℃ and in accordance with EN ISO 11607.
Sterilization of instruments by applying a fractionated
pre-vacuum steam sterilization process (according to EN
285/EN 13060/EN ISO 17665) under consideration of
the respective country requirements.
Minimum requirements: 5 min at 134 ° C.
Maximum sterilization temperature: 137° C.
Drying time: at least 8min.
Flash sterilization is not allowed on lumen instruments!
WARNING
 Use only approved autoclave devices according to
EN 13060 or EN 285.
 Respect
the
autoclave device given by the manufacturer.
 Use only this recommended sterilization procedure.
 Control the efficiency (packaging integrity, no
humidity, color change of sterilization indicators,
physicochemical integrators, digital records of cycles
parameters).
 The sterilization procedure must comply with EN ISO
17665.
 Waiting for cooling before touching.
Storage of sterilized instruments in a dry, clean and dust
free environment at modest temperatures, refer to label
testing
maintenance
procedure
according
to
of
the
Page 24 / 36