Biotage Selekt Installation And Safety
Hide thumbs Also See for Selekt:

Advertisement

Quick Links

Biotage
Selekt
®
Installation and Safety

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Selekt and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Biotage Selekt

  • Page 1 Biotage Selekt ® Installation and Safety...
  • Page 2: Table Of Contents

    Labels Used on the System Safety Requirements Restriction of Hazardous Substances (RoHS) Directive WEEE Compliance Statement Safety in Other Languages System Components General Information Consumables and Accessories Manufacturer Contact Us Biotage Selekt Installation and Safety | © Biotage 2018 ®...
  • Page 3: Installation

    Warning Site Requirements » The system must be unpacked and installed by an authorized Before the system is installed by an authorized Biotage service Biotage service engineer. engineer, the installation site should be prepared as follows: » Follow regional safety practices when handling and moving shipping boxes and containers, and when moving the system.
  • Page 4: Move The System

    If moving the system within the laboratory or between laboratories in the same building, follow the instructions below. If you need to ship the system, please contact Biotage 1-Point Support for instructions. 1. When the system is not processing, shut it down: a.
  • Page 5: Connections

    To prevent leakage, inspect all tubes before each run to ensure » Use only a power cord supplied by Biotage. The cord should be that they do not show signs of wear or damage and that they are inspected periodically and replaced if damaged.
  • Page 6: Technical Specifications

    1024 x 768 resolution; solvent stable. Ports: Ethernet port, USB ports (4), D-Sub 9-pin ports (2), and AUX port. See page 3. Certification CE and FCC marked and CB and NRTL certified. Biotage Selekt Installation and Safety | © Biotage 2018 ®...
  • Page 7: Safety

    Static Electricity The system must be operated in a fume hood that has the When using Selekt flash purification systems, care must be capacity to handle leakage of solvents and provide suitable taken to avoid static electricity build-up. This is particularly airflow to prevent the build-up of flammable vapors.
  • Page 8: Service

    Biotage service engineer. Before handing over the system for operating the system. Failure to install or use the system in the service, it should be emptied of liquid and cleaned from manner specified by Biotage may result in personal injury and/ harmful residues. or equipment damage.
  • Page 9: Restriction Of Hazardous Substances (Rohs) Directive

    (P/N 416183). Do not wipe away any excess solvent » from the column surface as this can generate static charge. Use only a power cord supplied by Biotage. The cord » should be inspected periodically and replaced if damaged.
  • Page 10: Safety In Other Languages

    Collection tray Drain port on the optional secondary solvent containment Secondary solvent containment (optional) Cooling fan - internal detector Touch screen Fuse holder and power inlet Figure 5. System components. Biotage Selekt Installation and Safety | © Biotage 2018 ®...
  • Page 11: General Information

    General Information General Information Consumables and Accessories Only genuine Biotage consumables and accessories must be used in the system. To order consumables and accessories, see contact information on the back of this document or visit our website www.biotage.com. Manufacturer Biotage Sweden AB...
  • Page 12 Part Number: 416182-A © 2018 Biotage. All rights reserved. No material may be reproduced or published without the written permission of Biotage. Information in this document is subject to change without notice and does not represent any commitment from Biotage. E&OE.