0.2 mL, and tubes with a 1.0 mL stem concentrate Figure 2. TurboVap LV manifold with 48 nozzles. In the picture to the right, to 0.7 mL. They all have a ±5% calibrated mark on the tube for 24 nozzles are plugged for usage with the multi rack with 24 positions.
Figure 4 . TurboVap LV multi racks with 48 positions (left) and 24 positions (right). Figure 7 . Gas vortex shearing action. Figure 5 . TurboVap II rack with (left) and without (right) end-point sensors. Ventilation Solvent vapors are removed by an exhaust fan and routed to the exhaust port at the rear of the system.
System Overview Sleep Mode and Automatic Wake-Up The system can be put in sleep mode either manually or by scheduling a sleep time. It will also enter sleep mode if the lid is left open for more than 90 minutes. When entering sleep mode, the system turns off the heater and the water bath light (if on), and activates the screen saver.
24 nozzles are plugged on the manifold (see Figure 2 on status field when the sample has been evaporated to dryness or page 1 on how to plug the TurboVap LV manifold), the to the desired liquid level. 24-position or Extrahera 1 x 24 manifold option must be selected in the software's settings view (see page 6).
» The desired end-point volume is greater than 0.2 or 0.7 mL. » The two end-point detection modes are only enabled when a TurboVap II end-point rack has been connected to the system's AUX port. System Ready Evaporate Settings 12:44 Wed 23 Nov 2016...
Define System Settings Calibrate the Water Bath Temperature The water bath temperature can be adjusted after calibrating at two temperatures that encompass (above and below) the range of temperatures used in your applications. 1. Set the evaporation mode to Manual or Time. 2.
1. Ensure that the water level is correct; see instructions above. 3. When the target water bath temperature has been reached, 2. If using the TurboVap LV multi rack with 24 positions or the the user will be notified by a dialog on the touch screen (see TurboVap EH sample/collection rack holder with only one Figure 18) and an audible alarm.
24 nozzles are plugged on the manifold (see Figure 2 sufficient reserve for the run. on page 1 on how to plug the TurboVap LV manifold), 3. If using a rack with end-point sensors, gently lower the rack...
Evaporate Monitor an Evaporation Run An audible alarm will also sound (but only once) each time the end-point level (+ any additional time) is reached for a single tube. During the evaporation run, the current and set values for the water bath temperature, gas flow, and time can be monitored on Note: The alarm and water bath light can be turned off in the the touch screen;...
» The power cord should be inspected periodically and replaced if by pressing on the top of the DRAIN port (see Figure 25B). damaged or altered. Use only a power cord supplied by Biotage. d. Empty the waste container. »...
Replace the Fuse Warning Figure 28. Releasing a vial spring. » Use only exact replacement fuses supplied by Biotage. Incorrect fuses create a potential fire hazard and personal injury. Replace a Nozzle 1. Turn off the system and disconnect the power cord.
4. Replace the old fuse. 5. Put the fuse holder back in place by pushing it in and turning it clockwise until it locks. If the fuse blows shortly after replacing it, please contact Biotage 1-Point Support ® ™ Figure 31. Removing the fuse holder.
If there is no or slow evaporation in a single tube: » » When using a system with TurboVap II configuration, If the end-point modes are disabled in the ensure that the nozzles for all rack positions that software, ensure that an end-point rack has contain sample tubes are enabled in the software.
General Information General Information Accessories and Spare Parts Part No. Description Only genuine Biotage accessories must be used in the system. C47811 10 mL Conical Centrifuge Test Tube To order consumables and accessories, see contact information on C44941 15 mL Conical Centrifuge Test Tube the back of this document or visit our website www.biotage.com.
Need help?
Do you have a question about the TurboVap and is the answer not in the manual?
Questions and answers