Selekt Preventive
®
Maintenance Checklist
Scope
The tasks carried out during the maintenance visit shall include
servicing the listed components and a final check which is
based on factory acceptance and qualifications tests. These two
procedures constitute a Biotage
(PM) Plan.
Each Preventive Maintenance visit should be completed within
two hours (per system) and the system returned to operational
status – unless of course a significant problem is identified.
During the PM visit, the system is disassembled, serviced,
re-assembled and given a performance / function test to help
ensure that it will remain in good working order for the following
year.
Preventive Maintenance is a recommended procedure by
Biotage to reduce the instrument downtime and assure reliable
operation, functionality and accuracy of your Selekt system.
Customer Responsibility
The customer should ensure that the site is prepared in agree-
ment with the guidelines provided in advance.
The checks that follow include the listed configurations
»
SEL-2SV, SEL-2SW, SEL-2EV, SEL-2EW
Customer Information
Company:
Name:
Email:
Phone:
© Biotage 2019
Selekt Preventive Maintenance
®
Items for the Technician to Bring
P/N
Description
Biotage
Selekt use PM kit P/N: 416832SP
®
416424SP, SFÄR-SAMPLET-
416832SP
TM1, FSUD-0443-0010,
416015SP, 413245SP
Spare Parts to Carry at PM Visit
SFAR-TM1
SFÄR SAMPLE TEST KIT #1
SFAR-TM1-
SFÄR SAMPLE TEST KIT #2
HMN
415115SP
INLINE CHECK VALVE
415950SP
ACCESSORY KIT
415599SP
DEUTERIUM LAMP
UV CELL, WITH 1/4-28
416331SP
THREAD
COLUMN HOLDER SFÄR 5G
416427SP
W. CLIP, 1 PCS.
413774
PTFE spray
416025SP
VALVE SOLVENT
416026SP
VALVE, HIGH PRESSURE
BIOTAGE
SELEKT PM
®
UI423
CHECKLIST
Test Tools to Carry at PM Visit
356633SP
CALIBRATION KIT
CALIBRATION ACCESSORY
416889SP
TOOL KIT
416411SP
CALIBRATION TOOL
System Information
System Serial#:
System model#:
Date of last PM:
Estimate date of next PM:
Biotage
Selekt PM Checklist | Page 1
®
Qty.
Notes
Grease, test mix
kit, Tube Column
1
Inlet CPL, FC
needle
Normal Phase
6
Chomatography
Reverse Phase
6
Chomatography
4
As needed
1
As needed
1
consumbale
1
As needed
1
As needed
1
Lubrication
1
As needed
1
As needed
1
Copy to carry
1
Pressure test tool
Pressure test tool
1
accessory
Fraction collector
1
calibration pin

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Selekt and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Biotage Selekt

  • Page 1 PTFE spray Lubrication Preventive Maintenance is a recommended procedure by 416025SP VALVE SOLVENT As needed Biotage to reduce the instrument downtime and assure reliable 416026SP VALVE, HIGH PRESSURE As needed operation, functionality and accuracy of your Selekt system. BIOTAGE SELEKT PM ®...
  • Page 2 If clogged, advise the customer to use sonication on the filters or how to order Inspect the solvent inlet filters. and replace with new. Refer to the user manual and the Maintenance chapter. Biotage recommendation is to replace filters every 1000 hours run time. Inspect and test the leak detectors. If applicable.
  • Page 3: Fraction Collector

    Biotage Selekt PM Checklist | Page 3 ® √ PM Task Notes Lubricate the pump screws. Refer to the Service manual for details and instructions. Replace the tube (column inlet) Part number 416015SP. Replace broken capillary holders if worn or broken. Part number 416427SP.
  • Page 4: Final Checks

    Verify in Service mode that all four fraction corner positions (both X and Y position) Run a Fraction Function test are correct and aligned over the proper spot. If needed calibrate using the Selekt calibration tool. Record the result. UV Detector...
  • Page 5: Record Section

    Literature Number: UI423 © 2019 Biotage. All rights reserved. No material may be reproduced or published without the written permission of Biotage. Information in this document is subject to change without notice and does not represent any commitment from Biotage. E&OE.