Specifications Portamess
Ranges
pH:
mV:
°C:
Display
LC display 35 x 67 mm, character height 15 mm
Measurement
approx. 1 s
Cycle
Measurement Error
pH:
(± 1 count)
mV:
°C:
Input
DIN 19 262
12
W
Input Resistance
> 1 x 10
-12
Input Current
< 1 x 10
A
(20 °C)
Electrode Stan-
Calimatic
automatic calibration with automatic buf-
dardization
fer recognition (German patent 29 37 227)
manual electrode standardization
Meter and Elec-
Sensoface
:
trode Monitoring
Meter self test:
Temperature Com-
Pt 1000 / NTC 30 kW
pensation
(automatic recognition during power-on) or manual
Data Retention
configuration and calibration data, factory settings
> 10 years
Auto Switch-off
after 1 hour
EMC
Emitted interference:
EN 61 326 Class B
Immunity to interference:
EN 61 326, EN 61 326/A1 and NAMUR NE 21
Environmental
Operation:
Temperature
Transport and storage:
Power Supply
3 alkaline AA cells
Operating Time
approx. 2,000 h*
* Due to storage, the service life of the included battery may be shorter.
911 pH
-2.00 to +16.00
-1,300 to +1,300
-20.0 to +120.0
< 0.01
< 0.1 % of meas. value ± 0.3 mV
< 0.3 K
evaluates the calibration interval,
zero point, electrode slope, re-
sponse time and glass impedance
of the electrode, optical indication
good/average/poor
during power-on
-10 to +55 °C
-20 to +70 °C
Appendix 25