Stopping a Training Session

To stop a training session both the DBWS feature and the Trolley Tracking feature must be turned
off, both buttons need to be green. See Figure 5-9. If the Previous Screen Button (See Figure 5-9)
is pressed, this will also end a training session. Once a training session is ended the fall limit setting
and dynamic unloading setting will go back to the default value.

Emergency Stop and Emergency Lower Buttons

At the bottom of the training session screen are the Emergency Stop and Emergency Lower buttons.
See Figure 5-9. At any given time these buttons can be used for emergency situations.
Emergency Stop Button
When the Emergency Stop Button is pressed the trolley stops moving and is in a locked position. The
system is in Static Unload mode.
Emergency Lower Button
If a patient becomes medically unstable the Emergency Lower Button can be pressed. Once pressed
the system will continue to winch down the rope, with controlled velocity, until the patient has reached
the ground or until the dynamic unload force goes below the default value.
Figure 5-10: Fall Prevented Window
Chapter 5: Patient Training Menu