Fall Limit Detection - Bioness vector ELITE Clinicians Manual

Body weight support system
Hide thumbs Also See for vector ELITE:
Table of Contents

Advertisement

Note: It is suggested that at the start of each exercise that the save button is pressed to ensure that
the zero-point of the fall limit value has been readjusted.
Previous Screen

Fall Limit Detection

During a training session the Vector Elite System will detect and prevent a fall when the fall distance
from the zero point is greater than the set Fall Limit value. When the system detects and prevents a
fall the trolley is locked and placed into Static Unload mode.
1. Vector Elite System detects a fall and the Fall Prevented window appears in the software. See
Figure 5-10.
2. Press the Manual Controls winch up button to help the patient regain control, so they are in
a standing position. Or if necessary use the Emergency Lower button to lower the patient to the
ground.
3. Close the Fall Prevented window and if the patient is in a standing position press the Dynamic
Body Weight Support button and (if applicable) the Trolley Tracking button to resume the
training session.
Clinician's Guide
20
Emergency Stop
Figure 5-9: Training Session Screen
Emergency Lower
Session Summary
Screen