Problem
During operation, no liquid
flows out from the insert tip.
poor performance
Table 9 – Troubleshooting Quick Guide
Possible Cause
the insert tip is of the type
which does not provide the
passage of liquids.
the water function is not
active.
the insert tip is clogged.
the handpiece is clogged.
the insert tip is not correctly
tightened on the handpiece
insert tip broken, worn or
deformed
TROUBLESHOOTING
Solution
Use an insert tip of the type
with the passage of liquids.
activate the water function.
Unscrew the insert tip from
the handpiece and unclog
the insert tip water passage
by blowing compressed air
through it. if the problem
persists, replace the insert tip
with a new one.
contact an authorized
mectron Service center
Unscrew and correctly screw
the insert tip back by using
the mectron torque wrench
chapter 4.4 on page
(see
12
of the Use and
Maintenance Manual)
Replace the insert tip with a
new one.