compact piezo LeD
WARNING: Infections control.
to ensure maximum safety of the patient and
the operator, before using all the reusable
parts and accessories, make sure you have
first cleaned and sterilized them, following
the instructions of the
17
of the Use and maintenance manual.
CAUTION: Contraindications. allow
reusable, autoclavable items (the handpiece,
the insert tips, the torque wrench, and any
other accessory that can be sterilized ) to
gradually return to room temperature after
steam sterilization and prior to usage. the
cooling process must not be accelerated.
WARNING: Breakage and wear-out of
the insert tips. High frequency oscillations
and wear-out may, in rare circumstances, lead
to the breakage of the insert tip.
Deformed or otherwise damaged insert tips
are susceptible to breakage during their use.
these insert tips must never be used.
if an insert tip breaks, check that none of its
fragments remain in the treated part and,
at the same time, apply effective suction to
remove them.
the patient must be instructed to breathe
through his nose during the treatment, or
a dental dam must be used to prevent the
patient from ingesting fragments of broken
insert tips.
check the level of wear of the insert tip and
its integrity before and during each use. if
there is a decrease in performance, replace
it. the level of wear of the most common
insert tips (S1, S1-S, S2, S5, P2, P4, P10) can
be checked by using the iNSeRt tip - caRD
supplied. to use the iNSeRt tip - caRD
correctly:
• position the insert tip on the iNSeRt
TIP - CARD so that the profile matches
the one printed on the card. The profile
printed on the card has a red line
indicating the wear limit;
• if the insert tip has a length less than
the wear limit, its performance will be
significantly lower than the condition
of a new one and replacement is
recommended.
4
Chapter 6 on page
if the layer of titanium nitride (golden
surface), where present, is visibly worn, the
insert tip must be replaced. the use of a worn
insert tip decreases its efficiency.
When the nitride coating wears out, the
cutting efficiency decreases; re-sharpening
the insert tip damages it and is therefore
forbidden. check that the insert tip is not
worn out.
During the intervention, frequently check that
the insert tip is intact, especially in its apical
part.
During the intervention, avoid prolonged
contact with retractors or with metallic
instrumentation in use. Do not exert excessive
pressure on the insert tips during their use.
WARNING: Only use original MECTRON
insert tips, accessories, and spare parts.
CAUTION: No modification of this
equipment is allowed.
WARNING: prior to any use, system
components must be inspected for damage.
Do not use if damage is apparent.
WARNING: Do not operate the device if
handpiece is defective, damaged or broken.
immediately replace the handpiece.
NOTE: in case of accident attributable to the
device during correct use and in accordance
with the intended use, a report must be
made to the competent authority and
manufacturer indicated on the product label.