Automatic NIBP double check

If the NIBP value exceeds the alarm limits, a new measurement takes place automatically
(immediately, when Manual measurement is selected, and after 30 seconds when Auto
measurement is selected.) If the alarm situation persists, an alarm is given.

Functioning of NIBP measurement during arrhythmias

The functioning of the NIBP measurement has been tested by using Biotek BP Pump W (Non-
invasive blood pressure monitor tester with wrist cuff) in the presence of the following
arrhythmias (available from the Biotek BP Pump W): Tachycardia, Bradycardia, Premature
Atrial Contraction #1, Premature Atrial Contraction #2, Premature Ventricular Contraction and
Atrial Fibrillation and PVCs.
The test results indicate that the presence of some arrhythmias during measurement may
increase the time required for the measurement.

Checklist

Check that:
The hose is correct: black NIBP hose for adults and children and light blue for infants.
The O-ring on the hose connector is intact.
The connector is firmly pushed inside the cuff tube.
The NIBP hose is properly connected to the module and will not detach if pulled.
The NIBP cuff is correct for the patient size.
There are no holes or cracks in the cuff bladder or cuff tube.
The symbol indicating the center of the bladder is over the artery.
All residual air is squeezed out of the cuff before wrapping it around the arm.
The cuff is not loose.
The cuff is at heart level.
The cuff tubes or NIBP tube are not kinked or squeezed together.
If leak is suspected, start venous stasis and check that the pressure is stable during stasis.
Non-invasive blood pressure is selected to be displayed through
Screen Setup.
Non-invasive blood pressure
Monitor Setup
-
16-9