Performance; Power Supply; Battery Operation; Environmental Conditions - GE Healthcare B30 User's Reference Manual

Patient monitor
Hide thumbs Also See for Healthcare B30:
Table of Contents

Advertisement

B30 Patient Monitor

Performance

WARNING
Operation of the monitor outside the specified values may cause inaccurate
results.
Any fluctuations within the specified limits do not affect the performance.
NOTE: Information in this section can be especially useful to clinicians.

Power supply

Rated voltages and frequencies:
Allowed voltage fluctuations:
Maximum power consumption:

Battery operation

Batteries:
Charging time:
Operation time:

Environmental conditions

Operating temperature:
normal operation:
while charging batteries:
Storage and transport temperature:
Relative humidity:
Atmospheric pressure:

Alarm behavior

The maximum alarm delay of the alarm at the monitor signal output to network: 5 seconds
If the alarm mode is latched, the technical alarms are latched as well. This does not comply with the NIBP
(IEC 60601-2-30) and invasive pressure (IEC 60601-2-34) standard requirements.
Silencing alarms for 5 minutes does not comply with the SpO
Defibrillator & IABP synchronization connector
Analog output
ECG:
From first user lead (ECG1)
Gain:
Delay:
DC offset:
Frequency response:
Invasive blood pressure:
From pressure labeled 'Art'
Gain:
Delay:
DC offset:
Frequency response:
The pacemaker pulses have been replaced with 2 ms ±20% fixed digital pulses at the ECG analog output.
A device that fulfils the requirements of the IEC 60601-1 standard can be connected to the defibrillator &
IABP synchronization connector. There are no other limitations, because the signals of the connector are
galvanically isolated from patient applied part of the ECG and invasive blood pressure measurements.
2-30
100 to 240 V 50/60 Hz
±10 %
150 VA
Exchangeable lithium-ion, 2 pcs max.
2 hours per battery pack
up to 4.5 hours
+5 to +40°C (41 to 104°F)
+5 to +35°C (41 to 95°F)
-20 to +60°C (-4 to 140°F)
10 to 90 % noncondensing
670 to 1060 mbar (500 to 800 mmHg)
(ISO 9919) standard requirements.
2
1 V/mV ±10%
< 15 ms
±100 mV max.
0.05 Hz to 40 Hz
10 mV/mmHg ±2%
< 35 ms
±20 mV max.
DC to 30 Hz

Advertisement

Table of Contents
loading

Table of Contents