B30 Patient Monitor
Table 10-8
Symptom
NIBP
measurement
does not work or
values seem
unstable
Table 10-9
Symptom
Temperature
measurement
fails
Table 10-10 Other problems related to SpO
Symptom
SpO
2
poor
Other situations
The following table lists some other situations that may occur during monitoring and possible
explanations.
Table 10-11 Other operation problems
Symptom
Printing is not
possible
The measured
values are not
displayed
The monitor does
not start
You cannot
perform a
measurement or
a function
10-8
Other problems related to NIBP measurement
Possible cause and solution
•
Check that cuff tubings are not bent, stretched, compressed or loose.
•
Prevent motion artifacts.
•
Use cuffs of correct size.
Other problems related to temperature measurement
Possible cause and solution
•
Check that you are using a correct probe.
•
Try another probe.
Possible cause and solution
•
signal is
Check the sensor and sensor positioning.
•
Change the averaging time from slow to normal.
•
Note that skin pigment causes differences.
•
Make sure that the patient is not shivering.
Possible cause and solution
•
Printer selection is None; change it through
connection.
•
Printer is not connected to the network. Check printer cable.
•
Check that you have selected the desired parameter to a waveform
or digit field.
•
Check that the power cord is properly connected.
•
Check that all cables are properly connected.
•
Check the fuses and replace them if necessary.
•
Check that the measurement module is properly installed.
•
Check that all cables are properly connected.
•
Remove the module and reinstall it.
measurement
2
Print/Record
- Printer