Other Situations - GE Healthcare B30 User's Reference Manual

Patient monitor
Hide thumbs Also See for Healthcare B30:
Table of Contents

Advertisement

B30 Patient Monitor
Table 10-8
Symptom
NIBP
measurement
does not work or
values seem
unstable
Table 10-9
Symptom
Temperature
measurement
fails
Table 10-10 Other problems related to SpO
Symptom
SpO
2
poor

Other situations

The following table lists some other situations that may occur during monitoring and possible
explanations.
Table 10-11 Other operation problems
Symptom
Printing is not
possible
The measured
values are not
displayed
The monitor does
not start
You cannot
perform a
measurement or
a function
10-8
Other problems related to NIBP measurement
Possible cause and solution
Check that cuff tubings are not bent, stretched, compressed or loose.
Prevent motion artifacts.
Use cuffs of correct size.
Other problems related to temperature measurement
Possible cause and solution
Check that you are using a correct probe.
Try another probe.
Possible cause and solution
signal is
Check the sensor and sensor positioning.
Change the averaging time from slow to normal.
Note that skin pigment causes differences.
Make sure that the patient is not shivering.
Possible cause and solution
Printer selection is None; change it through
connection.
Printer is not connected to the network. Check printer cable.
Check that you have selected the desired parameter to a waveform
or digit field.
Check that the power cord is properly connected.
Check that all cables are properly connected.
Check the fuses and replace them if necessary.
Check that the measurement module is properly installed.
Check that all cables are properly connected.
Remove the module and reinstall it.
measurement
2
Print/Record
- Printer