Optical Data; Microscope Carrier; Accessories; Floor Stand - Leica M525 F20/CT20 User Manual

Table of Contents

Advertisement

Technical Data

13.4 Optical data

Total magnification:
With 10× eyepiece: 1.2× to 12.8×
Field of view (Ø mm)
With 10× eyepiece: 16.5× to 180×

13.5 Microscope carrier

Turning range:
Rotation ± 335°
Lateral swiveling ±150° (right/left)
Tilting –30° to +120° (forward/
backward)
Balancing:
Motorized via A/B and C sledge
Brakes:
1 mechanical brake for
A/B axis
1 mechanical brake for C axis

13.6 Accessories

Binocular tube:
Variable viewing angle 30°–150°; built-in
interpupillary distance adjustment
permanently set to 45°
Beam splitter:
Two side outlets with light distribution of
50 % to 50 % or 70 % to 30 %
Video zoom tube:
3:1 zoom; focal distances of
35–100 mm; with C-mount;
individual focusing option
Second observer tube:
180° assistant's attachment with light
distribution of 70 % to 30 % or stereo
attachment for second observer to beam
splitter

13.7 Floor stand

Use:
Carries and positions the
Leica M525 F20/CT20 surgical microscope
Brakes:
3 mechanical brakes
Balancing:
Manual balancing at microscope carrier
and at swing arm
Handles:
Zoom adjustment
Working distance adjustment for
multifocal lens
Load:
Max. 6.5 kg from the interface of Leica
M525 F20/CT20
surgical microscope dovetail ring
Transport height:
1918 mm
Weight:
Approx. 229 kg with optics carrier and
handles
52
Leica M525 F20/CT20 / Ref. 10 715 156 / Version 01

13.8 Standards fulfilled

• Medical Devices Directive 93/42/EEC including amendments.
• Classification: Class I, in compliance with Annex IX, Rule 1 and
Rule 12 of the Directive.
• Medical Electrical Equipment, Part 1: General Requirements for
Safety in IEC 60601-1; EN 60601-1; UL 60601-1; CAN/CSA-C22.2
NO. 601.1-M90.
• Electromagnetic compatibility IEC 60601-2; EN 60601-1-2.
• Leica Microsystems (Schweiz) AG, Medical Division holds a
management system, which corresponds to the requirements of
the ISO 9001, ISO 13485 and ISO 14001 international standards
relating to quality management, quality assurance and
environmental management.

13.9 Environmental conditions

Use:
+10 °C to +40 °C
+50 °F to +104 °F
30 % to 75 % relative humidity
780 mbar to 1013 mbar air pressure
Storage:
–40 °C to +70 °C
–40 °F to +158 °F
10 % to 100 % relative humidity
500 mbar to 1060 mbar air pressure
Transport:
–40 °C to +70 °C
–40 °F to +158 °F
10 % to 100 % relative humidity
500 mbar to 1060 mbar air pressure