7. Press the H2 key and enter your measured value in the pop up window
8. Confirm this value by pressing the Enter key
9. Press the ESC key to save your stroke height calibration.
10. Return to the main page and then press the single key to go to the individual functions page
11. Lower the lift fork completely and record the height shown on the display as H3
12. Now raise the lifting fork completely and note the height displayed on the screen as H4
13. Now check whether H4 corresponds to the set maximum height (1.70m as default)
14. Check if H3 corresponds to the height you have noted down
15. Repeat the calibration if this is not the case.
Calibration battery
Max
1
Button to calibrate the battery
Do not press this button unless authorized to do so by the dealer or manufacturer.
Battery calibration should only be performed when the battery has a problem
Min
2
Battery calibration button
Do not press this button if you have not received approval from the dealer or manufacturer to do so.
The battery calibration should only be performed if the battery has a problem.
Lift Height Calibration
3
Press this button to enter the stroke height calibration and save the calibration of the battery.
ESC
4
Press this key to return to the settings page or to save your entries
Calibration of the battery
1. Open the control unit of the wheel gripper 2
® Copyright ATH-Heinl GmbH & CO.KG, All rights reserved / Misprints and technical changes reserved
34
Release date: 10-2022 / Product manufacturer ATH-Heinl GmbH & CO. KG