Keithley 6512 Instruction Manual page 72

Programmable electrometer
Table of Contents

Advertisement

MODEL NUMBER PREFIX
FUNCTION
0=VOLTS
1=AMPS
2=OHMS
3=COULOMBS
4=XFDBK
RANGE
Volts
Amps
Ohms Coulombs XFDBK
00 =
Auto
Auto
Auto
01 = 200mV
2 pA
2kΩ
02 =
2 V
20 pA
20kΩ
03 =
20 V
200 pA
200kΩ
04 = 200 V
2nA
2MΩ
05 = 200 V
20 nA
20MΩ
06 = 200 V
200 nA 200MΩ
07 = 200 V
2µA
2GΩ
08 = 200 V
20 µA
20GΩ
09 = 200 V
200 µA 200GΩ
10 = 200 V
2mA 200GΩ
11 = 200 V
20mA 200GΩ
12 = Auto off for all functions
ZERO CHECK
0 = OFF
1 = ON
ZERO CORRECT
0 = OFF
1 = ON
SUPPRESS
0 = OFF
1 = ON
TRIGGER
0 = CONTINUOUS, TALK
1 = ONE-SHOT, TALK
2 = CONTINUOUS, GET
3 = ONE-SHOT, GET
4 = CONTINUOUS, EXTERNAL
5 = ONE-SHOT, X
6 = CONTINUOUS, EXTERNAL
7 = ONE-SHOT, EXTERNAL
Figure 3-3
U0 status word and default values
DEFAULT:
0
00
1
6512 F
RR
C
Auto
Auto
200pC
200mV
2nC
2 V
20nC
20 V
20nC
20 V
20nC
20 V
20nC
20 V
20nC
20 V
20nC
20 V
20nC
20 V
20nC
20 V
20nC
20 V
0
0
6
0
0
0
0
Z
N
T
0
B
G
0
IEEE-488 Programming
7
00
0
= :
Q
MM
K
YY <CR LF>
TERMINATOR
ASCII
= : CR LF
: = LF CR
EOI; BUS HOLD-OFF
0 = 1 EOI + HOLD-OFF
1 = NO EOI + HOLD-OFF
2 = EOI + NO HOLD-OFF
3 = NO EOI + NO HOLD-OFF
SRQ
00 = DISABLED
01 = READING OVERFLOW
02 = DATA STORE FULL
08 = READING DONE
16 = READY FOR COMMAND
32 = ERROR
DATA STORE
0 = CONVERSION RATE
1 = 1 RDG/SEC
2 = 1 RDG/10 SEC
3 = 1 RDG/MIN
4 = 1 RDG/10 MIN
5 = 1 RDG/HR
6 = TRIG BUTTON
7 = DISABLED
ALWAYS 0
DATA PREFIX
0 = PREFIX, NO SUFFIX
1 = NO PREFIX OR SUFFIX
2 = PREFIX AND SUFFIX (IF B1
READING MODE
0 = ELECTROMETER
1 = DATA STORE
2 = MAXIMUM
3 = MINIMUM
ALWAYS 0
3-17

Advertisement

Table of Contents
loading

Table of Contents