Keithley 6512 Instruction Manual page 125

Programmable electrometer
Table of Contents

Advertisement

Maintenance
Table 7-13
Ranging amplifier checks
Step
Item/component
1
Range
2
Apply 190V DC to input
3
Range
4
Apply 19V DC to input
5
Range
6
Apply 1.9V DC to input
7
Range
8
Apply 190mV DC to input
Table 7-14
Digital circuitry checks
Step
Item/component
1
Range
2
U109, pin 40
3
U109, pin 2
4
U109, pin 3
5
U109, pin 4
6
U109, pin 6
7
U109, pin 37
8
U102, pin 6
9
U102, pin 3
Table 7-15
Display board checks
Step
Item/component
1
Power
2
U201, pin 9
3
U201,U202 pins 11-18
4
U204, pins 11-18, U205,
pins 15-18
5
U209, pins 10-15
6
P1016, pin 14
7
P1016, pin 15
8
P1016, pin 16
7-16
Required condition
Power-on, 200V DC range, zero check off
1.9V
Select 20V DC range
1.9V
Select 2V DC range
1.9V
Select 200mV DC range
1.9V
Required condition
Power-on, 200V DC range
+5V, ±5%
500µsec negative pulse every 1.6msec
Stays low <1sec on power-on, then goes
high (≈ 4V)
655kHz clock
655kHz clock
Variable pulse train
Variable pulse train
Negative-going pulse every 350msec
Required condition
Power-on, 200V DC range, zero check on
+5V, ±5%
Variable pulses
Variable pulses
1msec negative-going pulses every 10msec
Press COUL, ZERO CHECK, TRIG, or
DATA ON button
Press AMPS, SHIFT, RECALL, or SUP-
PRESS
Press VOLTS, OHMS, AUTO, ZERO COR-
RECT, or PROGRAM button
Remarks
Measure between 2V ANALOG
OUTPUT and COM.
Check X0.01 gain.
Check X0.1 gain.
Check X1 gain.
Check X10 gain.
Remarks
All voltages referenced to digital
common.
+5V digital supply.
Interrupt clock.
MPU reset.
Data strobe.
Address strobe.
A/D data.
A/D control.
Meter complete pulse
Remarks
All voltages referenced to digital
common.
+5V digital supply.
Segment drivers.
Digit drivers.
Pulse present when key is pressed.
Pulse present when key is pressed.
Pulse present when key is pressed.