Table 7-8
Power supply checks
Step
Item/component
1
S102 Line Switch
2
F101 Line Fuse
3
Line power
4
+5V Digital Supply
5
±210V Supplies
6
+5V Analog Supply
7
-9.1V Supply
8
±5V Bootstrap Supplies
9
±24 Supplies
Table 7-9
Relay configuration
Function
Range
K309
Volts
200mV
2V
20V
200V
Ohms
2kΩ
X
20kΩ
X
200kΩ
X
2MΩ
X
20MΩ
200MΩ
2GΩ
20GΩ
200GΩ
Amps
2pA
20pA
200pA
2nA
20nA
200nA
2µA
X
20µA
X
200µA
X
2mA
20mA
Coulombs
200pC
X
2nC
X
20nC
X
External
200mV
Feedback
2V
20V
Zero
*
Check
Required condition
Set to 115 or 230 as required.
Check for continuity.
Plugged into live receptacle; power on.
5V, ±5%
+240V, -240V, ±15%
+5V, ±5%
-9.1V, ±5%
+5V, -5V, ±5%
+27.5V, -27.5V, ±15%
K310
K308
K311
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
*
*
*
Relay
K312
K307
K303
K304
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
*
X
*
Maintenance
Remarks
Line voltage selection.
Remove fuse to check.
Referenced to digital common.
Referenced to analog common.
Referenced to analog common.
Referenced to analog common.
Referenced to bootstrap common.
Referenced to preamp out.
K305
K306
K301
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
*
K302
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
*
7-13