compact piezo p2K
Dry the insert tips torque wrench using a
soft cloth with low fibre release;
Lubricate the insert tips torque wrench
with
FDA approved medical-grade
lubricants by spraying it directly onto
the peripheral contact surface inside the
torque wrench, as indicated in the figure.
CAUTION: Do not use oil or silicone-
based lubricants.
after having applied the lubricant,
remove any excess oil using a clean lint-
free cloth.
6.4
Sterilization
The following parts and accessories require sterilisation:
• Scaler handpiece;
• Scaler handpiece nose cone;
• inser tips;
• Torque wrench.
6.4.1
Preparation
The packaging for terminally sterilized
instrument
should
requirements:
• Compliance with AAMI / ISO 11607-1;
• Legally marketed in the uS;
• Suitable for steam sterilization;
• Grade appropriate for weight of
instrument.
28
meet
the
following
3
4
WARNING: Diamond coated insert tips
are SINGLE USE ONLY. the diamond coated
insert tips are intended to be used on an
individual patient during a single treatment
procedure and then disposed of. the
diamond coated insert tips must be sterilized
only one time, prior the first use.
WARNING: ensure that the inner pouch
is large enough to contain the instruments
without stressing the seals or tearing the
packaging.