DISASSEMBLING O F P ARTS F OR C LEANING A ND S TERILIzATION
pushing it deeper. Do not apply pressure
to the insert in the axial direction.
• excessive pressure applied to insert tip
may cause insert tip fracture which may
cause harm to the patient and user.
Bending and/or prying an insert tip
reduces its structural integrity and can
result in an insert tip breakage during
use.
• Breakage and wear-out of the insert tips.
High frequency oscillations and wear-
out may, in rare circumstances, lead to
the breakage of the insert tip. Deformed
5
DISASSEMBLING OF PARTS FOR CLEANING
AND STERILIZATION
Before moving on to the cleaning procedures
described
in
Chapter 6 on page 17
disconnect all the accessories and components
of the compact piezo P2K.
if present, unscrew the insert tip from the
handpiece by using the torque wrench;
Disconnect the handpiece from its cord;
or otherwise damaged insert tips are
susceptible to breakage during their use.
these insert tips must never be used.
Should an insert tip fracture, during use,
check that none of its fragments remain
in the treated part and, at the same
time, apply effective suction to remove
them. the patient must be instructed
to breathe through his nose during the
treatment, or a dental dam must be used
to prevent the patient from ingesting
fragments of broken insert tips.
WARNING: Switch the device off.
,
Always turn the device off via the switch,
and disconnect the power supply cable
from the power outlet and from the device
body, before performing the cleaning and
sterilization interventions.
1
2
15