compact piezo p2K
thoroughly rinse the front cone under
running warm tap water and then under
demineralized water to eliminate any
residual detergent;
Hold the handpiece with its front end
pointed downward;
Rinse thoroughly the front end and the
surfaces of the external body of the
handpiece under running warm tap water
to eliminate any residual detergent;
FINAL RINSING
Hold the handpiece with its front end
pointed downward;
Rinse thoroughly the front end and the
surfaces of the external body of the
handpiece under demineralized water.
NOTE: Repeat steps 3 and 4 until rinse
de-mineralized water runs clear.
6.1.4
Insert Tips
place the insert tip in a clean container,
horizontally. Add enough of the enzymatic
detergent solution to completely cover it.
Soak the insert tip in the enzymatic
detergent solution for 10 minutes at
room temperature (22 ° C ± 2 ° C);
20
4
5
Need help?
Do you have a question about the compact piezo P2K and is the answer not in the manual?