Required Materials; Preparation - mectron compact piezo P2K Use And Maintenance Manual

Hide thumbs Also See for compact piezo P2K:
Table of Contents

Advertisement

compact piezo p2K
6.1.1
Required Materials
• Enzymatic detergent at pH 6-9;
• Water;
• container for immersion in the
enzymatic liquid;
• ultrasonic tank;
• Clean, soft cloths with low fibre release;
• Brush with soft nylon bristles;
• Syringe;
• Demineralised water.
6.1.2

Preparation

CAUTION: always disconnect the insert
from the handpiece before cleaning and
sterilising it.
Prepare an enzymatic detergent solution
with pH 6-9, as per the instructions of the
manufacturer;
CAUTION: once used, dispose of the
enzymatic detergent correctly, do not
recycle it.
18
CAUTION: To disinfect the device and/
or its accessories, it is recommended the use
of water-based disinfectant solutions with
aneutral pH (pH7).
DO NOT USE as disinfectants:
• Alcohol-based disinfectant solutions;
• Peracetic acid, Formaldehyde,
• Gluteraldehyde based products or other
equivalent solutions/products;
• very alkaline products (pH > 9);
• products containing sodium
hypochlorite;
• products containing hydrogen peroxide;
• products containing abrasive substances;
• very acid products (PH < 4);
• products containing aldehyde, amine
and/or phenols;
• acetone;
• Methylethylketone;
since they may discolour and/or damage the
plastic materials.
the manufacturer disclaims all liability for any
damage caused by the substances mentioned
above. in case of damage caused by those
substances, the Warranty will be void.