2.14 Down going pipeline over 5 m /16 ft length

Install air vent downstream of the flowmeter to prevent vacuum. Although this will not harm the
meter, it may cause gases to come out of solution (cavitate) and interfere with proper
measurements.
Figure 2-14: Down going pipeline over 5 m / 16 ft length
1
5 m / 16 ft
2 Install air vent

2.15 Mounting position

Figure 2-15: Allowed mounting position.
1 Partially filled is not applicable. The pipe line must always be fully filled
It is not allowed to install the device in a position, which is rotated in any degree wrt. the pipeline, other than horizontal.
2 Top - down position not applicable due to the buildup of gas and/or solids content.