Open Feed Or Discharge; Position Of Pump; Control Valve - KROHNE OPTISONIC 1400 Quick Start Manual

Ultrasonic insertion flowmeter for liquids
Table of Contents

Advertisement

OPTISONIC 1400

2.11 Open feed or discharge

Figure 2-11: Open discharge
Install meter on a lowered section of the pipe to ensure a full pipe condition through the meter.

2.12 Position of pump

CAUTION!
Never install flowmeter at a pump suction side in order to avoid cavitation or flashing in the
flowmeter.
Figure 2-12: Position of pump
1 For 3 paths:

2.13 Control valve

Figure 2-13: Installation in front of a control valve
3 DN
1
10/2016 - 4005626301 - QS OPTISONIC 1400 R01 en
15 DN, for 1 path or 2 paths: 30 DN
www.krohne.com