5
Replacing Assemblies
Removing and replacing the top panel and bottom panel
1
2
Figure 1
3
4
Figure 2
5
88
Disconnect the power cable, all probes and BNC input cables, and any other
cables including those for a mouse, keyboard, printer, USB, and LAN.
Remove the four M3 screws from the rear edge of the top panel. When
replacing, use a Torx T10 to torque to 5 in-lbs.
Slide the top panel back 1 to 2 inches and then lift it off.
To remove the bottom panel, flip the oscilloscope upside down and remove the
three M3 screws from the rear of the bottom panel.
Remove the remaining 13 screws from the bottom panel and lift the bottom
panel off the chassis.
Infiniium V-Series Oscilloscopes Service Guide