Initializing The Balance - AND GX-M Series Instruction Manual

Multi-function balance
Table of Contents

Advertisement

10-2 Initializing The Balance

This function returns the following parameters to factory settings.
□ Calibration data
□ Function table
The sample unit mass value (counting mode),
100% reference mass value (percent mode)
□ The data that is stored in the balance using the data memory function
□ External calibration weight
□ Function switch settings
□ Density of a liquid and a water temperature at density mode
Note
Be sure to calibrate the balance after initialization.
Setting procedure
1. Press the
ON:OFF
2. While pressing and holding the PRINT key and the SAMPLE
key, press the ON:OFF key to display p5 .
3. Press the
SAMPLE
4. Press the
PRINT
To cancel this operation, press the
5. Press the
RE-ZERO
6. With displaying
balance. The balance will automatically return to the weighing
mode.
key to turn off the display.
Clr
key to display
key.
CAL
key to change
Clr ,
press the PRINT key to initialize the
.
key.
.