Storage temperature
Climate class
Installation height as per
IEC61010-1 Ed.3
Degree of protection
Vibration resistance
Electromagnetic
compatibility (EMC)
34
30
316L housing material; process temperature range: –40 to +200 °C (–40 to +392 °F)
P1
= T
: –40 °C (–40 °F) | T
p
P2
= T
: +77 °C (+171 °F) | T
p
P3
= T
: +200 °C (+392 °F) | T
p
P4
= T
: +200 °C (+392 °F) | T
p
P5
= T
: –40 °C (–40 °F) | T
p
• Without LCD display: –40 to +90 °C (–40 to +194 °F)
• With LCD display: –40 to +85 °C (–40 to +185 °F)
DIN EN 60068-2-38 (test Z/AD)
• Generally up to 2 000 m (6 600 ft) above sea level
• Over 2 000 m (6 600 ft) under the following conditions:
• Supply voltage < 35 V
DC
• Power supply, overvoltage category 1
Test as per IEC 60529 and NEMA 250-2014
Housing and process connection
IP66/68, TYPE 4X/6P
(IP68: (1.83 mH
O for 24 h))
2
Cable entries
• Gland M20, plastic, IP66/68 TYPE 4X/6P
• Gland M20, brass nickel plated, IP66/68 TYPE 4X/6P
• Gland M20, 316L, IP66/68 TYPE 4X/6P
• Thread M20, IP66/68 TYPE 4X/6P
• Thread G1/2, IP66/68 TYPE 4X/6P
If the G1/2 thread is selected, the device is delivered with an M20 thread as standard and a G1/2
adapter is included with the delivery, along with the corresponding documentation
• Thread NPT1/2, IP66/68 TYPE 4X/6P
• HAN7D plug, 90 degrees, IP65 NEMA Type 4X
• M12 plug
• When housing is closed and connecting cable is plugged in: IP66/67 NEMA Type 4X
• When housing is open or connecting cable is not plugged in: IP20, NEMA Type 1
NOTICE
Plug M12 and plug HAN7D: incorrect mounting can invalidate the IP protection class!
‣
The degree of protection only applies if the connecting cable used is plugged in and screwed
tight.
‣
The degree of protection only applies if the connecting cable used is specified according to IP67
NEMA Type 4X.
‣
The IP protection classes are only maintained if the dummy cap is used or the cable is connected.
DIN EN 60068-2-64 / IEC 60068-2-64 for 5 to 2 000 Hz: 1.5 (m/s
• Electromagnetic compatibility as per EN 61326 series and NAMUR recommendation EMC (NE21)
• With regard to the safety function (SIL), the requirements of EN 61326-3-x are satisfied
• Maximum measured error during EMC testing: < 0.5 % of the span.
For more details refer to the EU Declaration of Conformity.
T
a
P1
P2
P5
: +77 °C (+171 °F)
a
: +77 °C (+171 °F)
a
: +38 °C (+100 °F)
a
: –40 °C (–40 °F)
a
: –40 °C (–40 °F)
a
Micropilot FMR60B HART
P3
T
p
P4
2
)
2
/Hz
Endress+Hauser
A0032024
Need help?