Reassembly - GE MAC 5500 Service Manual

Resting ecg analysis system
Hide thumbs Also See for MAC 5500:
Table of Contents

Advertisement

Reassembly

5-10
Maintenance: Disassembly Guidelines
3.
Turn the unit right side up and press the internal access button and raise the top
of the unit.
4.
Remove four (4) TORX screws.
Four (4) TORX screws
5.
Lower the top of the unit and lock in place.
6.
Raise the display to the vertical position.
7.
Gently lift the rear of the top cover free from the unit.
NOTE
The top cover holds the bezel that surrounds the rear panel connectors, so
the bezel may fall free at this time.
8.
At the front of the top cover, gently pull the thin strip of plastic free from under
the keyboard. The entire top assembly is now loose.
NOTE
It may be helpful to rotate the top cover 45
clear the display.
9.
Carefully lift the top assembly up and clear of the raised display.
1.
Raise the display to the vertical position.
2.
Make sure the bezel surrounding the rear panel connectors is in place.
3.
Lower the top cover down around the display and set in position.
4.
Snap the rear of the top cover in place and then, gently pulling on the thin
plastic strip at the front of the top cover, position it in place under the keyboard
assembly.
5.
Replace the screws removed in disassembly.
MAC 5500 resting ECG analysis system
2020299-020
to provide a larger opening to
Revision E