Point Of Use (Pre-Cleaning); Containment And Transportation - Hitachi EUP-VV531 Instruction Manual

Transvaginal probe
Table of Contents

Advertisement

 
The flowchart of the reprocessing process of this probe is as follows.   
Point of use 
(Pre‐cleaning) 
Containment and 
transportation 
 
4.1
Point of use (Pre‐cleaning)   
Pre‐cleaning should be done immediately after each use. The procedure is as follows: 
1) Remove the protective cover. 
2) Clean the probe of all patient's blood or fluid with running tap water until the surface 
of the probe looks visually clean. 
3) Wipe  the  whole  surface  of  the  probe  with  gauze  pad  and  remove  superficial  visible 
impurities. 
 
4.2
Containment and transportation 
Putting the contaminated equipment into exclusive shock and damage proof container for 
transportation  is  recommended.  It  is  recommended  that  instruments  are  reprocessed  as 
soon as possible and not later than 4 hours after usage.     
 
 
Manual cleaning and 
disinfection 
Manual Cleaning
Rinsing after manual 
cleaning 
Manual Disinfection
Rinsing after manual 
disinfection 
‐5‐ 
Drying 
 
Point of use 
(Pre‐cleaning) 
Containment and 
transportation 
Q1E‐EP1028