Hitachi EUP-VV731 Instruction Manual
Hitachi EUP-VV731 Instruction Manual

Hitachi EUP-VV731 Instruction Manual

Transvaginal probe

Advertisement

Quick Links

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Notes for operators and responsible maintenance personnel 
★  Please read through this Instruction Manual carefully prior to use. 
★  Keep this Instruction Manual together with the system with care to 
make it available anytime. 
 
 
.
 
Transvaginal Probe 
EUP‐VV731 
INSTRUCTION MANUAL 
Tokyo, Japan 
© Hitachi, Ltd. 2013, 2017 All rights reserved. 
 
Q1E‐EP1167‐6 
 
0123

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the EUP-VV731 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Hitachi EUP-VV731

  • Page 1                 Transvaginal Probe      EUP‐VV731      INSTRUCTION MANUAL                      Notes for operators and responsible maintenance personnel  ★  Please read through this Instruction Manual carefully prior to use.  ★  Keep this Instruction Manual together with the system with care to  make it available anytime.        Tokyo, Japan  Q1E‐EP1167‐6    © Hitachi, Ltd. 2013, 2017 All rights reserved.  0123  ...
  • Page 2     Manufacturer:  Hitachi, Ltd.  2‐16‐1, Higashi‐Ueno, Taito‐ku, Tokyo, 110‐0015, Japan  +81‐3‐6284‐3668    http://www.hitachi.com/businesses/healthcare/index.html      European    Representative:  Hitachi Medical Systems GmbH  Otto‐von‐Guericke‐Ring 3 D‐65205 Wiesbaden, Germany        EU Importer:  Hitachi Medical Systems Europe Holding AG  Address:  Sumpfstrasse 13 CH‐6300 Zug, Switzerland      Local Distributor:      (1)  Q1E‐EP1167  ...
  • Page 3   About this manual  This instruction manual shall provide instructions for using, cleaning, and disinfecting the  HITACHI  ultrasound  probes.  It  also  describes  safety  considerations,  maintenance.    For  instructions  for  operating  the  main  unit,  refer  to  the  operation  manual  for  it.    Before  using the probe, thoroughly read this manual and keep this book for future reference.  If you have any questions concerning the manual, please contact a service support.    The  following  conventions  are  used  throughout  the  manual  to  denote  information  of  special ...
  • Page 4   Graphical Symbols for Use in Labeling of Hitachi Ultrasound Probes  Some  graphical  symbols  that  are  used  in  labeling  of  Hitachi  Ultrasound  Probes  are  compliant with EN980:2008 standard. Refer to the following table about the meanings of  them.    Explanation of Symbol Symbol  Descriptive Content  Hitachi,Ltd  Manufacturer Company  2‐16‐1, Higashi‐Ueno,taito‐ku,  Name and Address    Tokyo,110‐0015,Japan  Authorized  Hitachi Medical Systems GmbH  Representative in The  Otto‐von‐Guericke‐Ring 3 D‐65205 Wiesbaden,  European Community  Germany  Store the probe in a cool, dustproof and dark,  Keep away from  dry environment to avoid high temperature  Sunlight  and humidity and direct sunlight.     ...
  • Page 5   Definition of symbol  The following symbol is also used for HITACHI Ultrasound Probes.  Location  Symbol  Definition  Probe  This  instrument  complies  with  Directive  93/42/EEC  relating  to  connector  Medical Device and Directive 2011/65/EU relating to RoHS  Probe  IPX7 mark  IPX7  connector  See section 1.5.  Probe  Type BF APPLIED PART  connector    Probe  General warning sign  connector    Probe  Warning; dangerous voltage  connector    Probe  Caution; Biohazard  connector    Follow  the  instruction  manual  to  operate  this  instrument.  If  not ...
  • Page 6: Table Of Contents

          CONTENTS  Page Introduction ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 1 Features  ∙ ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 1 Principles of operation ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 1 Intended Use  ∙ ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 1 Composition  ∙ ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 2 Construction∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 2 Inspection before Use ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 3 Inspection for Appropriate Connection  ∙ ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 3 Inspection for Material Surface ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 3 Operation Procedure ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 4 Cleaning and Disinfection ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 5 Point of use (Pre‐cleaning)  ∙ ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 6 Containment and transportation ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 6 Manual Cleaning and disinfection ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 7 Drying ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 8 Inspection ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 8 Maintenance and Safety Inspection  ∙ ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 9 Daily check ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 9 Safety Precautions ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 10 Specifications ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 11 Probe ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 11 Suppliers List ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 12 Disposal of the probe  ∙ ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙ 12  ...
  • Page 7: Introduction

    3) The  transducer  is  also  used  to  receive  reflected  ultrasonic  waves.  The  transducer  vibrates  mechanically  due  to  the  received  ultrasonic  waves  and  converts  mechanical  vibrations into electric energy. Electric signals are converted to shades of brightness by  brightness modulation to obtain an image.    Intended Use  The Transvaginal Probe EUP‐VV731 is designed for observation and diagnosis mainly of  the following regions by connecting with the HITACHI ultrasound scanner.     Transvaginal     Transrectal         ...
  • Page 8: Composition

    1) Probe EUP‐VV731 ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙  1 piece  2) Instruction Manual∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙  1 copy  3) EUP‐VV731 Correct Backlash CD‐ROM  ∙ ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙  1 piece    CAUTION   1) Sterilization has not been made to the probe, shipped from the factory. Prior to use of  the probe, be sure to clean and disinfect the probe.  2) This  CD‐ROM  is  important  for  loading  specific  parameter  of  the  probe  to  the  ultrasound scanner by Hitachi engineer during the installation. Avoid missing it and  keep it certainly.    Construction  Immersible part (IPX7)  Un‐Immersible part 150mm from the tip of head Head Applied ...
  • Page 9: Inspection Before Use

      2. Inspection before Use  Prior to use, the probe must be carefully inspected that it is appropriate for use. If not, do  not use the probe and immediately contact a service support.      Inspection for Appropriate Connection  1) Check that the system is correctly operating.    Refer to the instruction manual for the  Ultrasound diagnostic scanner.  2) Never  attach  or  connect  unauthorized  devices  or  instruments  on  the  probe,  such  as  unauthorized biopsy attachments.    Inspection for Material Surface  1) Visually  check  the  surface  of  the  probe  head,  housing,  cable  and  connector  for  any  crack, scratch or denaturalization. ...
  • Page 10: Operation Procedure

      3. Operation Procedure  1) Confirm that the probe is cleaned and disinfected.  2) Connect  the  probe,  operate  the  Main  unit,  and  adjust the image according to the instructions given  in the operation manual for the main unit.  3) Relationship  between  direction  of  the  probe  and  Index mark  the  right‐left  orientation  mark  on  the  image  is  (Projection) shown in Fig. 2.  4) Before  a  diagnosis,  put  sterilized  acoustic  jelly  to  Right‐left ...
  • Page 11: Cleaning And Disinfection

      4. Cleaning and Disinfection    The probe and accessory must be reprocessed after each use. Refer to the  reprocessing instruction in this chapter.    ‐ The probe is delivered unsterile. Prior to the first use, reprocess the  probe.  WARNINGS  ‐ Temperature should not exceed 50°C during reprocessing.  ‐ Probe connector is not water resistant.  The probe is not completely submersible. The immersible part is  Limitations on  shown in Fig. 3. The un‐immersible part should be disinfected by wipe  reprocessing  disinfection.      Sterilization is not possible for this probe.  The probe should be packed in a sterile pouch or container to transport  Transportation  from Central Sterile Supply Department (CSSD) to an operating room.  before using  Be careful not to damage the sterile pouch or container during  transportation.  Levels of reprocessing requirements:    Depending on the application of the product and with regard to risk evaluation, the user  has to classify the medical device according to the current Medical Device Directive for  processing of medical devices as uncritical, semi‐critical or critical. Supporting information  concerning this topic is listed in the table below. The user is responsible for correct  classification of the medical device.    Classification  Definition  Processing  Cleaning ...
  • Page 12: Point Of Use (Pre-Cleaning)

      The flowchart of the reprocessing process of this probe is as follows.    Manual cleaning and  disinfection  Point of use  Manual Cleaning Drying  (Pre‐cleaning)  Rinsing after manual  cleaning  Containment and  transportation  Manual Disinfection Rinsing after manual  disinfection      Point of use  Point of use (Pre‐cleaning)    (Pre‐cleaning)  Pre‐cleaning should be done immediately after each use. The procedure is as follows:  1) Remove the protective cover.  2) Clean the probe of all patient’s blood or fluid with running tap water until the surface  of the probe looks visually clean.  3) Wipe  the  whole  surface  of  the  probe  with  gauze  pad  and  remove  superficial  visible  impurities. ...
  • Page 13: Manual Cleaning And Disinfection

      Manual Cleaning and disinfection    Prepare following items before manual cleaning and disinfection:    a) Detergent: Cidezyme (Johnson & Johnson, #2258) or another cleaning agent with  approved material compatibility for this medical device  b) Disinfectant: Cidex OPA (Johnson & Johnson, # 20391) or another disinfectant with  approved material compatibility for this medical device  c) Two tanks, one for cleaning and one for disinfection ‐ optional:  1 additional tank for rinsing with deionized/tap water (sufficient size for immersion of  the immersible part of the probe at full length)  d) Soft, fluff free cloth or single use towel  e) Personal protective equipment (gloves, water repellent protective skirt, face protection  mask or protective glasses, see also instructions of the manufacturer for the detergent  and the disinfectant)  Manual Cleaning:    Prepare the detergent solution in a tank with cold water (please follow the instructions of  the detergent manufacturer regarding application, dilution and contact time).  1) The temperature of the detergent solution should be between 15‐30°C, concentration is  1.6%.  Please  note  the  minimum  contact  time  of  the  detergent  in  the  manufacturer’s  instruction. ...
  • Page 14: Drying

      Manual disinfection:    1) Prepare  the  disinfectant  solution  in  a  tank  with  cold  water  (please  follow  the  instructions  of  the  disinfectant  manufacturer  regarding  application,  concentration,  microbiological efficiency, service life and contact time).  2) Confirm  the  concentration  of  the  disinfectant  before  immersing  the  probe.  Although  ® Cidex  OPA does not need to be diluted, it is recommended to use test strips to verify ...
  • Page 15: Maintenance And Safety Inspection

        5. Maintenance and Safety Inspection    Daily check  1) Visually  check  the  surface  of  the  probe  head,  housing,  cable  and  connector  for  any  crack,  scratch  or  denaturalization.  If  you  find  damage,  do  not  use  the  probe  and  immediately contact a service support.    2) After using the probe, it should be cleaned and disinfected according to “4. Cleaning  and  disinfection”, then  store  it  in a  cool and dark  place  avoid  high temperature  and  humidity, direct sunlight. ...
  • Page 16: Safety Precautions

      6. Safety Precautions  WARNING   1) If you find that the probe head, housing or cable are cracked or damaged, stop use of  the probe immediately. Please contact a service support.  2) Warning in case of using probe covers which latex is contained to. The latex may cause  such  allergic  reactions  as  itching,  rubor,  urticaria,  swelling,  fever,  anhelation,  wheezing, depression of blood pressure, shock and so on.    For the patients suspected  of  latex  allergy,  do  not  use  the  latex‐containing  medical  devices.    If  your  patient  shows  any  of  above  mentioned  symptoms  during  the  operation,  stop  the  use  of  the  latex‐containing ...
  • Page 17: Specifications

      7. Specifications  Probe    Type:  EUP‐VV731 Transvaginal probe  Center frequency:  5.5MHz  Technology:  Convex Array Probe  Dimensions:  See Fig. 4.  Weight:    Approx. 1.0kg (including cable and connector)  Scanning angle:    140°  Volume sweep angle:  80°  Probe materials:  Biocompatible allergy free components  Acoustic output:  According to IEC 60601‐2‐37 (See Main Unit manual)  Applicable systems:  Depending on production and upgrade status.    For  detailed information, contact a service support.  Classification:  MDD classification IIa.  Cleaning:  Applicable detergents are listed in the suppliers list  Disinfection:  Applicable disinfectants are listed in the suppliers list  Operating conditions:    Ambient temperature:  +15 to +40°C  Contact surface temperature ...
  • Page 18: Suppliers List

      Suppliers List  The products listed below are seriously tested and approved for use with the Transvaginal  Probe EUP‐VV731.    Product name  Manufacturer  Purpose  CIDEZYME®  Johnson&Johnson Cleaner  Cidex®  Johnson&Johnson Disinfectant Cidex® OPA  Johnson&Johnson Disinfectant PERASAFETM  DuPont  Disinfectant CIV‐FlexTM 610‐006  CIVCO  Probe cover CIV‐FlexTM 610‐007  CIVCO  Probe cover Please contact your local distributor for a current version of the ʺDisinfectant/Sterilization  Method Compatibility for Ultrasound Probe and Accessory List”.        8. Disposal of the probe  Recycle or dispose equipment properly in compliance with your organizational rules and  your local laws.      CAUTION   Before disposing equipment, disinfect or take other infection‐prevention measures.  Disposal ...
  • Page 19   Unit: mm  Fig. 4  Dimensions of EUP‐VV731      ‐13‐  Q1E‐EP1167  ...

Table of Contents