5.1 Point of use (Pre-cleaning)

Pre-cleaning should be done immediately after each use.
The procedure is as follows:
1) Remove any accessories from the probe, like probe covers, biopsy
attachment, magnetic sensor attachment or needle guide bracket.
2) Clean the probe of all patient's blood or fluid with running tap
water until the surface of the probe looks visually clean.
3) Wipe the whole surface of the probe with gauze pad and remove
superficial visible impurities.

5.2 Containment and transportation

Putting the contaminated equipment into exclusive
shock and damage proof container for transportation
is recommended. It is recommended that instruments are reprocessed as
soon as possible and not later than 4 hours after usage.

5.3 Manual Cleaning and disinfection

Prepare following items before manual cleaning
and disinfection:
a) Detergent: Cidezyme® (Johnson & Johnson,
#2258) or another cleaning agent with
approved material compatibility for this
medical device.
b) Disinfectant:
Johnson, # 20391) or another disinfectant
with approved material compatibility for
this medical device.
c) Two tanks, one for cleaning and one for
disinfection - optional:
1
additional
deionized/tap water (sufficient size for
immersion of the submergible part of the
probe at full length)
d) Soft, fluff free cloth or single use towel
e) Personal protective equipment (gloves, water repellent protective
skirt, face protection mask or protective glasses, see also
instructions of the manufacturer for the detergent and the
disinfectant)
Cidex®
OPA
(Johnson
tank
for
rinsing
-16-
(Pre-cleaning)
Containment and
transportation
Manual cleaning and
disinfection
Manual Cleaning
Rinsing after manual
&
cleaning
Manual Disinfection
with
Rinsing after manual
disinfection