Transmitting Results to the Host
SA and UA results are transmitted to the Host in two ways:
• If the Auto Send option is enabled in the Host configurations, the VETSCAN SA will
transmit the result to the Host when the result is saved to the SA's database at the end of a
sample run. The result will not be transmitted to the Host until the result is saved to the SA
Database. This rule applies to Retest workflows also.
• Results can be transmitted to the Host from the SA Database by selecting the result and
clicking the Send button.
Once bidirectional communication with a Host is enabled, results in the Database will display the
most resent transmission status of a result.
Send Status
Function
Auto
Result successfully, automatically sent to PIMS.
Fail
Result failed sending to PIMS. Network or EMR Error.
Sent
Result successfully sent to PIMS by user manually selecting result from the
SA Database and clicking the Send button.
Print
Result has been sent to printer. Result has been manually printed or Auto
Print is On with Auto Send Off.
Final
Result has been saved, but not sent to the PIMS and not printed. Auto Send
Off and Auto Print is off.
(blank)
Result has been retested with Retest workflow. Result has not been printed
and has not been sent to PIMS. If desired, select Print or Send.
PIMS – Practice Information Management System
EMR – Electronic Medical Record
VETSCAN SA Network/Host Connectivity Statuses
VETSCAN SA Network/Host Connectivity Statuses are only reported if Send by network port is
enabled in the Host configurations of the SA.
Type
Error No.
N/A
N/A
N/A
N/A
Warning
NW14
Warning
NW17
Warning
NW18
Description
Icon
Off
Connected
Ethernet
Cable Error
DHCP Error
EMR Error
Status Definition
Send by network port is disabled.
Connectivity between the Host and SA
is operational and ready for bidirectional
communication.
No connectivity with the Host because
there is no network cable connected to
the VETSCAN SA or the network cable
connected to the VETSCAN SA is not
connected to the network.
No connectivity with the Host because
VETSCAN SA was unable to get a valid IP
address from the network's DHCP server.
Connectivity between the FUSE and SA is
non-operational.
VE TS CA N SA US ER ' S M A N UA L
1 . 2 S OF T WA RE
|
4 1