Maintenance Schedule - Zoetis vetscan SA User Manual

Sediment analyzer
Table of Contents

Advertisement

Maintenance Schedule

Maintenance
Cleaner Reagent Cycle
Kit Change
Empty Waste Container
Clean Fan Filter
It is recommended to complete daily cleaning around the analyzer including wiping off any
sample spills. All VETSCAN SA external component parts should be kept clean. The outside and
inside of the sample probe will be automatically cleaned with Rinse A after each use. Rinse B will
be run automatically daily at midnight.
Do not use the following chemicals to clean VETSCAN SA parts. Exposure to these chemicals
may cause plastic parts to crack, craze, etch, or warp.
Alcohols
o Methanol
o Ethanol
o Isopropyl alcohol
Chlorinated hydrocarbons
o Chloroform
o Carbon tetrachloride
Aromatic hydrocarbons
o Benzene
o Phenol
o Acetone
Do not use abrasive or highly alkaline cleaners (example: undiluted bleach) on VETSCAN
SA parts. Do not expose VETSCAN SA parts to temperature > 140 °F (60 °C). Do not sterilize
VETSCAN SA parts by autoclaving or dry heat.
Frequency
Once per week
Depends on reagent usage
Depends on reagent usage
As needed
Software reminder
Yes. Prompts user to complete
Cleaner cycle before running a new
test.
Yes. Reagent kit or individual rinse
depleted warning. Additional time
based reminder warning when the
pack has been opened 53 days.
Yes. Reagent waste full warning.
No. Check periodically.