3. Click Run.
4. Mix urine sample by gently inverting.
Ensure a thoroughly mixed sample.
5. Immediately place tube below sample probe and press blue start paddle.
1. Hold the sample below the sample probe leaving
a gap to the probe tip to allow the sample probe to
aspirate air prior to moving down for sampling.
2. Ensure sample probe touches tube bottom. and
maintains contact until the probe starts to move
back up.
While measuring, the SA screen is locked and navigation to other screens is prohibited. Also,
the SA status indicator light (top right of the SA Results Panel) will display yellow during
measurement process. During a SA Retest of a SA+UA result, the completed UA results will
continue to show on the UA Results Panel.